LENSTAR LS 900® Technical Specs

The below mentioned measurement ranges are based on the standard settings of the device for automatic measurement and analysis.

Corneal Thickness(CT)

 

Kerometry

 

Measurement range

300 - 800 µm

Measurement range for radius

5 - 10.5mm

Display resolution

1µm

Display resolution

0.01mm

In-vivo repeatability

± 2µm

In-vivo repeatability

± 30µm

Anterior chamber depth(ACD)

 

Measurement range for axis angle

0-180°

Measurement range

1.5 - 5.5mm

Display resolution

Display resolution

0.01mm

In-vivo repeatability

± 9°

In-vivo repeatability

± 20µm

White-to-white distance

 

Lens Thickness

 

Measurement range

7 - 16mm

Measurement range

0.5 - 6.5mm

Display resolution

0.01mm

Display resolution

 

In-vivo repeatability

± 0.3mm

In-vivo repeatability

± 50µm

Pupillometry

 

Axial Length (AL)

 

Measurement range

2 - 13mm

Measurement range

14 - 32mm

Display resolution

0.01mm

Display resolution

0.01mm

Eccentricity of the visual optical line

 

In-vivo repeatability

± 25µm

Display resolution

0.01mm

 

 

Retinal Thickness

 

 

 

Manually Assessed

 

 

 

Display resolution